Spark Tester Calibrator
Spark Tester Calibrator
For Spark Testing with constant and comparable sensitivity on insulating, winding and sheathing lines, it is very essential to calibrate the spark tester with a spark tester calibrator.
Katakkar Engineering make Spark Tester Calibrator Model STCAL-2K is designed to meet the requirements of relevant Indian and international standards for Spark Testers. One can measure the test voltage and it’s frequency with the help of Spark Tester Calibrator.
Incorporating Digital Voltmeter, this versatile Calibrator is used for both Voltage Calibration and measurement of Fault detection sensitivity of various makes and type of spark testers.
Measurement and display circuits for :
- Test Voltages up to 15 kV RMS for AC High Frequency Spark Testers
- 30 kV RMS / DC for AC Mains Frequency / DC Spark Testers resp
- Test Voltage Frequency upto 4000 Hz
Type of measurements for test voltage and short circuit current:
- Average
- RMS
- Positive and Negative Peak Value
Artificial fault device
Checks the sensitivity to ‘faults’ according to IS:10810 (part 44) – 1984, BSEN 50356:2002
Technical Specifications :
General:
Parameter | Specifications |
---|---|
Mechanical Dimensions | 490 x 270 x 280 mm |
Total weight of Calibrator | 15 kg |
Input | 1 Phase, 230 VAC + 10%, 50 Hz + 2 Hz |
Power Rating | 300 VA |
Technical Specifications :
For measurements of test voltages
Parameter | Specifications |
---|---|
Input impedance | 100 M for AC mains frequency & DC Spark Testers 40 M for AC High Frequency Spark Testers. |
Max. Test Voltage | 15 kV at max. 4000 Hz for AC high frequency spark testers 30 kV RMS / DC for AC mains frequency / DC spark testers |
Technical Specifications :
For Artificial fault device
Parameter | Specifications |
---|---|
Max. operating voltage | 3 kV AC RMS for AC High frequency & Mains frequency Spark testers 5 kV DC for DC Spark testers |
Fault Rate | 1 per second with 50 Hz mains frequency |
Spark Gap | 0.25mm + 0.05 mm for a period of 0.0025 second once per second |
Max. fault current | 1 per second with 50 Hz mains frequency |
Input | 1 phase, 230 V AC + 10%, 50/60 Hz + 2 Hz |